For complete protection in all clinical situations, choose the correct level of Face Mask for your application. Performance tests have now changed to ensure that the level of protection against blood penetration is clearly understood.
Revised in December 2015 the standard AS 4381:2015 for Face Masks are now performance based on 3 tests:
Bacterial Filtration Efficiency (BFE) with Differential Pressure (Delta P) & Synthetic Blood Penetration Resistance. Fluid resistance is measured mm Hg, masks are rated according to performance.
L1-80 mm Hg. L2-120 mm Hg. L3-160 mm Hg.
Fluid Resistance: The ability of the mask to limit the passage of blood or body fluids from the outer facing through the mask layers to the inner facing.
Level 1. Applications:For general purpose medical procedures, where the wearer is not at risk of blood or body fluid splash or to protect staff and/or the patient from droplet exposure to microorganisms (e.g. patient with upper respiratory tract infection visits General Practitioner).
Level 2. Applications:For use in emergency departments, dentistry, changing dressings on small wounds or healing wounds where minimal blood droplet exposure may possibly occur (e.g. endoscopy procedures).
Level 3 Applications:For all surgical procedures, major trauma first aid or in any area where the health care worker is at risk of blood or body fluid splash (e.g. orthopaedic, cardiovascular procedures).
Med-Con Level 1, 2 & 3 Masks to AS 4381:2015: Replace fluid resistant masks. 4-layer mask with ultra-soft non-reactive inner layer to maximize facial comfort. All Nelson Laboratory tested to required methods.
Automated process of manufacturing ensures minimum bio-burden contamination. No wastage as total product is consistent, and guaranteed batch-to-batch regularity. Material selection combined with Ultrasonic Sealing provides long time comfort and performance for the user. Standard of manufacture ensures the highest level of quality and service for our customers.
Covered PVC Nose wire – Loops or Ties – Green or Blue (at request). Latex Free.
Surgical Face Masks to AS 4381:2002: 3-layer mask with ultra-soft non-reactive inner layer to maximize facial comfort. Bacterial protection for the wearer as well as the protection of the patient. Particulate protection in industrial applications. Med-Con masks test at a PFE of 99.1% with a particle size of 0.1um.Delta pressure of less than 1.5 units optimises users’ breathability.
Surgical Face Masks – AS 4381:2015
Single Use Face Masks For Use In Health Care
Level 1, 2 & 3
Masks to AS 4381:2015
Ties and Loops
Description
For complete protection in all clinical situations, choose the correct level of Face Mask for your application. Performance tests have now changed to ensure that the level of protection against blood penetration is clearly understood.
Revised in December 2015 the standard AS 4381:2015 for Face Masks are now performance based on 3 tests:
Bacterial Filtration Efficiency (BFE) with Differential Pressure (Delta P) & Synthetic Blood Penetration Resistance. Fluid resistance is measured mm Hg, masks are rated according to performance.
L1-80 mm Hg. L2-120 mm Hg. L3-160 mm Hg.
Fluid Resistance: The ability of the mask to limit the passage of blood or body fluids from the outer facing through the mask layers to the inner facing.
Level 1. Applications:For general purpose medical procedures, where the wearer is not at risk of blood or body fluid splash or to protect staff and/or the patient from droplet exposure to microorganisms (e.g. patient with upper respiratory tract infection visits General Practitioner).
Level 2. Applications:For use in emergency departments, dentistry, changing dressings on small wounds or healing wounds where minimal blood droplet exposure may possibly occur (e.g. endoscopy procedures).
Level 3 Applications:For all surgical procedures, major trauma first aid or in any area where the health care worker is at risk of blood or body fluid splash (e.g. orthopaedic, cardiovascular procedures).
Med-Con Level 1, 2 & 3 Masks to AS 4381:2015: Replace fluid resistant masks. 4-layer mask with ultra-soft non-reactive inner layer to maximize facial comfort. All Nelson Laboratory tested to required methods.
Automated process of manufacturing ensures minimum bio-burden contamination. No wastage as total product is consistent, and guaranteed batch-to-batch regularity. Material selection combined with Ultrasonic Sealing provides long time comfort and performance for the user. Standard of manufacture ensures the highest level of quality and service for our customers.
Surgical Face Masks to AS 4381:2002: 3-layer mask with ultra-soft non-reactive inner layer to maximize facial comfort. Bacterial protection for the wearer as well as the protection of the patient. Particulate protection in industrial applications. Med-Con masks test at a PFE of 99.1% with a particle size of 0.1um.Delta pressure of less than 1.5 units optimises users’ breathability.
Proudly made in Australia. ARTG No. 109872
Med-Con Level 1, 2 and 3 Face Mask 2018